Forumslovakia doping bust: 13 detained, 80+ kinds of peptides and two labs seized · Shop talk · 3 replies
Home › Testing Labs › Clark Turner Scientific

Clark Turner Scientific

Could not verify · checked

Peptigrity could not verify Clark Turner Scientific as a testing laboratory when it last checked, on Aug 7, 2026. Below is each check, what came back and the date, from Peptigrity's own posts in the thread about it.

What Peptigrity checked

13 checks, from Peptigrity's posts in the thread
CheckWhat we askedWhat came backDate
ReviewsReviews of the labNone found.
Contact emailanalysis@clarkturner.com, the address the site publishesIt can't receive email: the domain's owner has published an instruction not to deliver mail to it, and the domain is listed for sale at $3,495. The lab's own domain, clarkturnerscientific.com, has working email through Zoho.
Company recordA company under this name in registriesNo record found.
Address, accreditation and termsA physical address, an accreditation number and terms of service on the siteNone listed.
PeopleAny person associated with the labNone independently confirmed.
How the PDFs were madeThe creator recorded in the three certificate filesAll three record Chrome's "print to PDF", on a non-standard page size.
SignatureThe handwritten name at the bottomTyped text in a free Google font that looks like handwriting.
InstrumentThe instrument each spectrum namesThe header names one type of instrument; the settings table under it lists controls that exist only on a different type, from a different manufacturer. On all three.
Instrument settingsThe settings across the three runsIdentical on all three, for three substances tested two months apart.
Retention timeWhether the reported times fit the column and flow rate listedOne report has the substance coming out at 2.35 minutes, before the roughly 2.5 minutes the liquid itself takes.
Mass resultsWhether the weight measurements match the formula printed on the same reportThey agree with each other, not with the formula.
Batch numbersThe header's batch number against the data file shown on the same pageThey differ on two of the three.
Verification requestCompany name, address, accreditation status and the original instrument files, with 14 days to replyNo reply a couple of days in.

What we could verify

Three certificates are attributed to the lab: three compounds, three batch numbers, dated March and May 2026. Their chemical formulas are correct, the purity arithmetic is sound and one of the three spectra holds up. Peptigrity doesn't name the shop they came from (staff post, Aug 7, 2026). A member posted in the thread on Jul 23, 2026.

The thread about Clark Turner Scientific

clarkturnerscientific.com testing lab. - real or not? →

Peptigrity lists labs it could not verify so buyers can see what was checked and when. They are not in the lab directory's counts, its price table or any comparison. Labs we could not verify →